visit the hl7 website The Demo site for our new HL7 Version 2+ (plus) Standard

18.7.13 TCU/ACK - Automated Equipment Test Code Settings Update (Event U10) (13.3.10)

This message is used to send information concerning test codes and parameters from one application to another (e.g., automated equipment to a Laboratory Automation System). This message transfers the current snapshot of the test parameters of the sending system. The sent parameter sets are supposed to replace the parameter sets existing at the receiver of this message before the trigger (there is no selective "Add" or "Delete").

Segment Cardinality Implement Status
TCU^U10^TCU_U10
MSH

Message Header

[1..1] SHALL
SFT

Software Segment

 
EQU

Equipment Detail

[1..1] SHALL
TEST_CONFIGURATION [1..*] SHALL
SPM

Specimen

[0..1]  
TCC

Test Code Configuration

[1..*] SHALL
ROL

Role

[0..1]   B

 

TCU_U10

MSH-15 MSH-16 Immediate ACK Application Ack
Blank Blank - TCU^U10^TCU_U10
Blank Blank - ACK^U10^ACK
NE NE - -
NE NE - -
AL, SU, ER NE ACK^U11^ACK -
AL, SU, ER NE ACK^U10^ACK -
NE AL, SU, ER - TCU^U10^TCU_U10
NE AL, SU, ER - ACK^U10^ACK
AL, SU, ER AL, SU, ER ACK^U11^ACK TCU^U10^TCU_U10
AL, SU, ER AL, SU, ER ACK^U10^ACK ACK^U10^ACK
We need some ER7 examples...
We need some XML examples...